• Model / Type: Dual Channel Tourniquet Pump, REF 5920 011 000
• Software: Version 1.3.x
• Power Supply (input): 100–240 VAC, 50/60 Hz (universal power supply)
• AC Protection: Fuse 3.15 A / 250 V
• Current Consumption: approx. 1.5 A at 120 VAC
• Internal Power: 15 VDC ±1 %, max. 4 A, 60 W with protection
• Battery: NiMH, 12 VDC, 2.6–2.7 Ah, operating time up to 5 hours, charging 1 h (80%) / 6 h (full)
• Dimensions: approx. 43.2 × 25.4 × 22.9 cm
• Weight: approx. 5.4 kg
• Cuff Pressure Range: 100–475 mmHg (accuracy ±2% or ±4 mmHg)
• Timer: 1–240 minutes, adjustable every 1 min (up to 15 min) and every 5 min (above 15 min)
• Control: independent dual channel control
• Air Pump: 12 VDC, 1.25 A, flow rate approx. 10 L/min
• Pressure Transducers: range 0–15 psi, accuracy ±0.2%
• Alarms: power loss, overload, overpressure, system errors